Field Instrument Calibration Procedures - MCQs Quiz

Test Your Knowledge: Field Instrument Calibration

1. What is the primary purpose of calibrating a field instrument?

2. The documented readings of an instrument before any adjustments are made during calibration are known as:

3. In a 5-point calibration, what are the typical calibration points checked?

4. The difference between the highest and lowest value that an instrument can measure is called its:

5. A pressure transmitter has a calibrated range of 0 to 200 psi. What is its span?

6. Hysteresis error in an instrument is the difference in output reading for the same input when:

7. Which of the following is the first adjustment typically made during the calibration of a standard analog transmitter?

8. A "smart" transmitter with HART protocol offers which of the following advantages during calibration?

9. The gradual change in an instrument's measurement over time is known as:

10. What does the term "traceability" in calibration refer to?

11. For a standard 4-20 mA current loop, a reading of 12 mA corresponds to what percentage of the measured value?

12. When calibrating a differential pressure (DP) level transmitter on a closed tank, what is typically done to the low-pressure side?

13. Which of the following is a common tool used for calibrating pressure gauges in the field?

14. The "As Left" data in a calibration certificate represents:

15. A linearity error in an instrument means that:

16. The acceptable deviation of a measurement from the true value is known as:

17. For calibrating a thermocouple, the most accurate reference would be:

18. What is the purpose of a 3-way manifold on a differential pressure transmitter?

19. If a pressure transmitter with a range of 0-150 bar and a 4-20 mA output is reading 8 mA, what is the corresponding pressure?

20. Zero suppression in a transmitter is used when:

21. In a field calibration, what is the primary reason for checking the loop power supply?

22. Which part of a pH probe requires regular calibration?

23. A "turndown ratio" for a transmitter refers to:

24. Which of the following is a non-contact method for level measurement calibration?

25. Before performing a calibration on a device in a live process, what is the most critical first step?

26. The span adjustment on a transmitter primarily corrects for:

27. A flow meter is being calibrated. What would be a suitable standard for comparison?

28. An "Out of Tolerance" (OOT) condition found during calibration implies that:

29. What information is typically found on a calibration sticker?

30. The process of adjusting a controller to optimize its performance in response to process changes is known as:

Leave a Reply

Your email address will not be published. Required fields are marked *