Troubleshooting Pressure Instruments – Top 25 Interview Questions & Answers
1. What are the first steps you would take when a pressure transmitter is showing an erratic or fluctuating reading?
My initial approach is to systematically isolate the cause, starting from the process and moving toward the instrument.
- Verify the Process: First, I'd check if the process itself is unstable. I would consult with a process operator to confirm if there are known issues like pump cavitation, turbulent flow, or control valve hunting that could cause real pressure fluctuations.
- Check for Environmental Factors: I'd inspect the instrument for external influences like vibration from nearby machinery or rapid temperature changes, which can affect sensor output.
- Inspect Impulse Lines: The connection between the process and the transmitter is a common source of problems. I would check for blockages, leaks, trapped air (in liquid service), or condensate (in gas service) within the impulse tubing.
- Check Power Supply: I would verify that the transmitter has a stable power supply (typically 24 VDC) and that the wiring connections are secure and free of corrosion.
- Enable Damping: As a temporary measure to stabilize the reading for control purposes, I might increase the damping value on the transmitter's configuration, but I would continue to investigate the root cause.
2. A pressure transmitter is reading a constant zero (or minimum value). What are the potential causes?
A zero reading points to a few key areas, from a simple blockage to a complete failure.
- Isolation Valve Closed: The most common and easiest issue to check is whether the isolation valve (or root valve) at the process tap is closed, preventing any pressure from reaching the sensor.
- Plugged Impulse Line: The line connecting the transmitter to the process could be completely blocked by solidified process media, debris, or hydrate formation in cold weather.
- Transmitter Failure: The sensor diaphragm or internal electronics of the transmitter may have failed. This can be verified by applying a known pressure source directly to the transmitter with a calibrator.
- Loss of Power: A complete loss of power to the instrument will result in a zero reading (or a specific failure mode signal like 2.4 mA). I would check fuses, wiring, and the power source.
- Incorrect Ranging: While unlikely to cause a perfect zero, if the transmitter is ranged incorrectly (e.g., expecting a very high pressure when the actual pressure is low), it might read at the bottom of its scale.
3. How do you troubleshoot a pressure transmitter that reads consistently high or low?
This suggests a calibration issue or a systematic error in the installation.
- Zero/Span Shift: The most likely cause is a calibration drift. The instrument needs to be recalibrated. This involves isolating it from the process, venting to atmospheric pressure to check the zero point, and then applying a known pressure (typically the upper range value) with a calibrator to check the span.
- Static Head Pressure: If the transmitter is mounted above or below the process tapping point in a liquid service, the column of liquid in the impulse line creates a hydrostatic head pressure. This will cause a constant high (if below) or low (if above) reading unless it has been compensated for in the transmitter's calibration (zero elevation or suppression). I would verify the installation height against the calibration settings.
- Trapped Pressure: The instrument or impulse line might not have been fully vented before being returned to service, trapping pressure and causing a positive offset (reading high).
- Incorrect Compensation: For differential pressure (DP) flow or level measurements, incorrect density compensation values for the process fluid will lead to a consistently incorrect reading.
4. What is a "five-point calibration," and why is it important for pressure instruments?
A five-point calibration is a standard procedure to verify the accuracy and linearity of an instrument over its entire calibrated range.
- Procedure: It involves applying known, stable pressure inputs at five points: 0%, 25%, 50%, 75%, and 100% of the calibrated span. The output of the transmitter is recorded at each point, both in an upscale and downscale direction.
- Importance:
- Linearity Check: It confirms that the instrument's output is directly proportional to the input across the entire range. A simple two-point (zero and span) calibration doesn't check the points in between.
- Hysteresis Check: By checking both upscale and downscale, it reveals any hysteresis, which is the difference in output for the same input value depending on whether the pressure is increasing or decreasing.
- Repeatability: The test confirms that the instrument gives the same output for the same input consistently.
- "As Found" and "As Left": Technicians typically perform an "As Found" five-point check to document the instrument's state before any adjustments and an "As Left" check after calibration to prove it meets accuracy specifications.
5. How would you handle a leaking impulse line on a critical gas service?
Safety is the absolute priority when dealing with a leak, especially on a critical gas service which could be flammable, toxic, or high-pressure.
- Immediate Safety Actions: First, I would assess the risk. This includes identifying the gas, understanding its hazards (flammability, toxicity), and determining the leak severity. I would wear appropriate Personal Protective Equipment (PPE) and use a gas detector to monitor the area. If necessary, I would barricade the area and inform control room and area supervisors immediately.
- Isolate the Source: The safest way to stop the leak is to isolate the pressure source. This involves closing the root valve at the process tap. I would follow proper Lock-Out/Tag-Out (LOTO) procedures to ensure the valve is not reopened accidentally.
- Depressurize and Vent: Once isolated, the impulse line must be safely depressurized. This is done by opening the vent valve on the instrument manifold, directing the vented gas to a safe location away from personnel and ignition sources.
- Repair the Leak: With the line fully isolated and depressurized, I can safely repair the leak. This typically involves tightening fittings or replacing the damaged section of tubing or the faulty fitting.
- Reinstatement: After the repair, I would perform a leak test (e.g., using soapy water or a leak detection fluid) at operating pressure before fully returning the instrument to service.
6. Explain the difference between gauge pressure, absolute pressure, and differential pressure.
The key difference is the reference point from which the pressure is measured.
- Gauge Pressure (psig, barg): This is pressure measured relative to the local atmospheric pressure. A gauge pressure transmitter has a vent that allows the "low side" of its sensor to reference the ambient air pressure. A reading of 0 psig means the pressure is the same as the surrounding atmosphere. This is the most common type of pressure measurement.
- Absolute Pressure (psia, bara): This is pressure measured relative to a perfect vacuum (absolute zero pressure). An absolute pressure sensor has a sealed vacuum on its reference side. This measurement is independent of changes in atmospheric pressure and is critical for applications where atmospheric pressure variations would affect the process, such as boiling point control.
- Differential Pressure (psid, bard): This is simply the difference between two applied pressures. A DP transmitter has two process connections: a high-pressure (HP) port and a low-pressure (LP) port. It measures the difference (P1 - P2) and is used extensively for measuring fluid flow (across an orifice plate), liquid level (hydrostatic head), and filter blockages.
7. You need to calibrate a DP transmitter used for flow measurement. What information do you need before starting?
Calibrating a DP transmitter for flow requires more than just the pressure range; it requires understanding the entire flow loop.
- Primary Element Details: I need the data sheet for the primary flow element (e.g., orifice plate, venturi, or pitot tube). This will specify the differential pressure produced at a maximum flow rate (e.g., "250 inH2O @ 1000 GPM"). This DP value becomes the Upper Range Value (URV) for the transmitter.
- Process Fluid Density: The calculation for the DP range is based on a specific fluid density. I need to know the operating density of the fluid to ensure the calibration is accurate. If the density changes, the flow reading will be incorrect unless compensated for.
- Transmitter Range (URV/LRV): I need to know the Lower Range Value (LRV) and Upper Range Value (URV) that the transmitter should be configured for. For flow, the LRV is almost always 0. The URV is the max DP from the data sheet.
- Output Signal: I need to confirm the desired output, which is typically 4-20 mA. I also need to know if the relationship is linear or square root. For DP flow, a square root extraction function is usually applied in the DCS or the transmitter itself to provide a linear flow signal.
- Required Tools: A precision pressure calibrator (e.g., a HART communicator with a pressure module or a deadweight tester), appropriate fittings, and tools for connecting to the manifold.
8. What is the purpose of a 3-valve or 5-valve manifold on a DP transmitter?
A manifold is a critical accessory that allows for safe and efficient isolation, venting, and equalization of the transmitter without disrupting the main process line.
- 3-Valve Manifold:
- High-Pressure Block Valve: Isolates the high-pressure side of the process.
- Low-Pressure Block Valve: Isolates the low-pressure side of the process.
- Equalizing Valve: Connects the high and low-pressure sides together. This is crucial for checking the transmitter's zero point under static line pressure and preventing catastrophic over-range damage to the sensor during startup or shutdown.
- 5-Valve Manifold: Includes the three valves above plus:
- Two Vent/Test Valves: One for the high-pressure side and one for the low-pressure side. These allow for safe venting of trapped pressure and provide connection points for a calibrator to apply a test pressure. This is much safer and more convenient than cracking a fitting.
9. Describe how you would perform a "zero check" on a live DP transmitter.
A zero check under static pressure is a quick and effective way to verify the transmitter's health without a full calibration.
- Communicate: First, I would inform the control room operator that I am about to perform a check. The control loop may need to be put in manual mode to prevent process upsets.
- Isolate: I would close the low-pressure block valve first, then the high-pressure block valve on the manifold. This traps the process pressure inside the manifold and transmitter.
- Equalize: I would slowly open the equalizing valve. This applies the same trapped static pressure to both sides of the DP sensor. Since the differential pressure is now zero, the transmitter's output should drop to its 4 mA (or zero DP) reading.
- Verify: I would check the output using a HART communicator or a multimeter in series with the loop. If the reading is not exactly at the zero point, a zero trim adjustment may be necessary.
- Return to Service: To return to service, I perform the steps in reverse: close the equalizing valve, slowly open the high-pressure block valve, and then slowly open the low-pressure block valve. Finally, I would inform the operator that the check is complete.
10. What are some common issues related to heat tracing on impulse lines?
Heat tracing is used to prevent process media from freezing or solidifying in impulse lines, but it can cause its own set of problems if not properly designed or maintained.
- Overheating: If the heat trace temperature is too high or the controller fails, it can vaporize a liquid in the line. This creates gas pockets, which are compressible and will lead to erratic or incorrect pressure readings.
- Cold Spots/Failure: If the heat trace fails or has inconsistent coverage, cold spots can develop. This can lead to the fluid becoming more viscous or solidifying, causing a partial or full blockage of the line.
- Incorrect Installation: Heat trace tubing should not be in direct contact with the transmitter body itself, as this can overheat the electronics and cause premature failure or measurement drift.
- Insulation Issues: Damaged or wet insulation is ineffective and can lead to the heat trace being unable to keep the line at the required temperature, increasing the risk of freezing.
11. What is HART protocol and how is it useful in troubleshooting?
HART (Highway Addressable Remote Transducer) is a hybrid communication protocol that superimposes a digital signal on top of the standard 4-20 mA analog signal.
- How it Works: It allows for two-way communication with the smart instrument without interrupting the primary 4-20 mA control signal. This is done using a HART communicator (field communicator) or a modem connected to a laptop.
- Troubleshooting Benefits:
- Remote Diagnostics: You can see detailed diagnostic information from the device, such as "sensor failure," "electronics failure," "loop current fixed," etc. This tells you exactly what the transmitter thinks is wrong.
- View Multiple Variables: You can view not only the primary process variable (pressure) but also secondary variables like sensor temperature and the analog output value in milliamps. You can instantly see if the mA output matches the pressure reading.
- Configuration and Calibration: You can remotely check and change the device's configuration (range, damping, units) and perform trims (zero and span adjustments) without needing to apply physical pressure in some cases.
- Loop Testing: You can command the transmitter to output a specific mA value (e.g., 8 mA, 12 mA) to test the integrity of the entire loop, from the wiring back to the control system input card.
12. Explain the concept of "zero suppression" and "zero elevation."
These are calibration techniques used in DP level measurement to compensate for the mounting position of the transmitter relative to the vessel tappings.
- Zero Elevation (Wet Leg): This is used when a DP transmitter is mounted below the bottom process tap of a vessel. The column of liquid in the high-pressure impulse line exerts hydrostatic pressure on the sensor even when the tank is empty (0% level).
- Effect: This causes the transmitter to read high.
- Compensation: The transmitter's zero point is "elevated" to a positive pressure value. For example, if the head pressure from the liquid column is 50 inH2O, the LRV is set to +50 inH2O, which will correspond to a 4 mA output.
- Zero Suppression (Dry Leg): This term is often used interchangeably with zero elevation, but it specifically applies to closed/pressurized tank applications where the low-pressure side (reference leg) is intentionally filled with a liquid (a "wet leg") to provide a stable reference.
- Effect: The wet leg exerts a constant hydrostatic pressure on the LP side. The transmitter constantly "sees" a negative differential pressure.
- Compensation: The range is shifted or "suppressed" so that a specific negative DP value corresponds to the 4 mA output.
13. A pressure gauge and a pressure transmitter are installed on the same process tap, but they show different readings. How do you determine which one is correct?
This is a common scenario. I would not assume either is correct without verification.
- Isolate and Test: The most reliable method is to isolate both instruments from the process and connect them to a single, external pressure source like a hand pump with a high-accuracy digital calibrator.
- Apply Test Pressure: Apply a known pressure to all three devices (the gauge, the transmitter, and the calibrator) simultaneously. This will immediately show which instrument, if any, is out of calibration. The calibrator is the trusted reference.
- Consider Installation Differences: If an in-place test isn't possible, I'd check for installation issues. Is one device mounted higher than the other in a liquid service? A difference in elevation will cause a difference in hydrostatic head pressure, leading to different readings.
- Check for Blockages: The tapping point for one device could be partially plugged while the other is clear.
14. What is the purpose of a diaphragm seal (remote seal)? When would you use one?
A diaphragm seal is a device that allows a pressure instrument to measure the pressure of a process without the process fluid ever touching the sensor itself.
- How it Works: It consists of a flexible diaphragm connected to the instrument via a capillary tube filled with a stable, incompressible fluid (like silicone oil or glycerin). Process pressure pushes on the diaphragm, and this pressure is hydraulically transferred through the fill fluid to the instrument's sensor.
- When to Use One:
- Corrosive or Erosive Fluids: To protect the expensive transmitter sensor from fluids that would quickly damage it.
- High Temperature Processes: To move the transmitter away from extreme heat that would damage its electronics. The capillary tube allows it to be mounted in a cooler location.
- Viscous or Clogging Fluids: For processes like slurries, pulp, or heavy oils that would clog the small port of a standard transmitter. The large surface area of the diaphragm prevents this.
- Hygienic/Sanitary Applications: In food, beverage, and pharmaceutical industries, diaphragm seals provide a smooth, crevice-free surface that is easy to clean and prevents product contamination.
15. What are common problems associated with diaphragm seals?
While very useful, diaphragm seal systems introduce unique potential failure points.
- Fill Fluid Leakage: A leak in the capillary tube or at the diaphragm weld will cause the system to lose its fill fluid, leading to a loss of pressure transfer and an incorrect (usually low or zero) reading.
- Temperature Effects: The fill fluid will expand and contract with changes in ambient temperature. This can cause a zero shift in the reading. This is a significant issue for systems with long capillary tubes, though many modern systems have compensation features.
- Hydrogen Permeation: In certain processes, hydrogen atoms are small enough to pass through the metal diaphragm and form gas bubbles in the fill fluid. These bubbles are compressible and will cause measurement errors and drift. Gold plating the diaphragm can mitigate this.
- Damaged Diaphragm: The diaphragm itself can be dented or damaged by physical impact or overpressure, which will affect its ability to accurately transfer pressure, leading to calibration issues.
16. How does ambient temperature affect a pressure transmitter, and how is it compensated for?
Ambient temperature can affect both the sensor and the electronics, causing zero and span shifts.
- Effect on Sensor: Temperature changes can cause the physical components of the sensor (diaphragms, strain gauges, silicon wafers) to expand or contract, altering their physical properties and leading to measurement drift.
- Effect on Electronics: The performance of electronic components like resistors and amplifiers can also drift with temperature, affecting the signal conditioning and output.
- Compensation: Modern smart transmitters have built-in temperature sensors located near the primary pressure sensor. The transmitter's microprocessor contains a characterization map that knows how the sensor behaves at different temperatures. It constantly measures the internal temperature and applies a corresponding correction factor to the pressure reading in real-time, actively compensating for ambient temperature effects. This is a major advantage of smart transmitters over older analog ones.
17. You are checking a 4-20mA loop for a pressure transmitter and measure 22.5 mA. What does this indicate?
A reading above the normal operating maximum of 20 mA (typically 20.5-20.8 mA for over-range) is a specific diagnostic signal.
- Failure High Condition: This is a standard failure mode according to the NAMUR NE 43 recommendation. Most smart transmitters can be configured to drive their output to a high alarm state (e.g., ≥ 21.0 mA) or a low alarm state (e.g., ≤ 3.6 mA) to signal a critical internal fault.
- Potential Faults: The high reading indicates that the transmitter has detected a serious internal problem. This could be:
- A sensor diaphragm failure.
- An internal electronics or microprocessor failure.
- A failure in the analog-to-digital converter.
- Troubleshooting Steps: I would connect a HART communicator to the device. The communicator will provide a specific diagnostic message (e.g., "Sensor Failure") that explains why the transmitter has gone into this failure state. The instrument will likely need to be replaced.
18. What is "turndown ratio" and why is it important?
Turndown ratio, or rangeability, describes the flexibility of a transmitter's operating range.
- Definition: It is the ratio of the maximum possible calibrated span (the Upper Range Limit or URL) to the minimum possible calibrated span the instrument can be set to while still maintaining its specified accuracy.
- Example: A transmitter has a URL of 400 inches of water column (inH2O) and a turndown ratio of 100:1. This means you can accurately calibrate it for a minimum span of 400 / 100 = 4 inH2O. You could use this single transmitter for an application requiring a 0-400 inH2O range or for another application requiring only a 0-10 inH2O range.
- Importance:
- Inventory Reduction: Plants can stock a single model of a high-turndown transmitter and use it for a wide variety of applications, reducing the need for many different models.
- Application Flexibility: If process conditions change and a new range is needed, a high-turndown transmitter can simply be re-ranged without needing to be replaced.
- Performance Note: It's important to remember that accuracy is typically specified as a percentage of the *calibrated span*, not the URL. As you "turndown" a transmitter to a smaller span, the potential error becomes a larger percentage of the actual reading.
19. How do you select the correct pressure instrument for a given application?
Selecting the right instrument requires a thorough understanding of the process conditions.
- Pressure Range: The instrument's range must encompass the normal operating pressure as well as the maximum and minimum expected pressures. The normal operating point should ideally be between 25% and 75% of the calibrated span.
- Process Fluid Properties: I need to know the fluid's chemical composition (for material compatibility), viscosity, temperature, and whether it contains solids. This determines if a diaphragm seal is needed and what materials (e.g., 316 Stainless Steel, Hastelloy C, Monel) are required for the wetted parts.
- Accuracy Requirement: Critical applications may require a high-accuracy transmitter (e.g., 0.05% of span), while less critical monitoring points might be fine with a lower accuracy instrument (e.g., 0.5% of span).
- Environment: The instrument's housing must be suitable for the installation environment. This includes temperature rating, ingress protection (IP rating) for dust/water, and hazardous area classification (e.g., Intrinsically Safe or Explosion Proof).
- Output/Protocol: I need to ensure the instrument provides the required output signal (e.g., 4-20mA HART, Foundation Fieldbus, Profibus PA) compatible with the control system.
20. What is an Intrinsically Safe (I.S.) pressure transmitter?
Intrinsic Safety is a protection technique used for electronic equipment installed in hazardous areas where flammable gases or dust may be present.
- Principle: The principle of I.S. is to limit the amount of electrical energy (voltage and current) available in the instrument and its wiring to a level below what is required to ignite the most volatile mixture of a specific hazardous atmosphere.
- How it's Achieved: An I.S. system consists of three parts:
- The I.S. Field Device (the transmitter), which is designed to not store significant energy.
- The I.S. Barrier or Isolator, located in the safe area, which is the key component that limits the energy sent into the hazardous area.
- The Interconnecting Wiring, which has specific capacitance and inductance limits.
- Benefit: The major advantage of I.S. is that it allows for live maintenance and calibration in the hazardous area without needing to de-energize the circuit or obtain a "hot work" permit, which significantly improves efficiency.
21. A pressure switch is failing to actuate at its setpoint. What steps would you take?
Troubleshooting a pressure switch involves verifying both the pressure it's seeing and its mechanical/electrical function.
- Isolate and Vent: First, I would safely isolate the switch from the process and vent any trapped pressure.
- Connect a Calibrator: I would connect a variable pressure source (hand pump) with a reference gauge to the switch's input port.
- Check Electrical Continuity: I would connect a multimeter (in continuity or resistance mode) across the switch contacts (Common and Normally Open/Normally Closed).
- Test Actuation Point: I would slowly increase the pressure from the calibrator and watch the multimeter to see at what pressure the contacts change state (actuate). I would note this "trip" pressure.
- Test Deactuation Point: I would then slowly decrease the pressure and note the pressure at which the contacts reset. The difference between the trip and reset points is the "deadband" or "hysteresis."
- Compare and Adjust: I would compare the measured trip point to the required setpoint. If they don't match, I would use the switch's setpoint adjustment screw to change it, re-testing until it actuates at the correct pressure. If it cannot be adjusted correctly or operates erratically, the switch is faulty and must be replaced.
22. What is the overpressure limit of a transmitter and why is it important?
The overpressure limit is a critical safety and performance specification of any pressure instrument.
- Definition: It is the maximum pressure that can be applied to the instrument without causing permanent damage to the sensing element. It is typically specified as a multiple of the instrument's Upper Range Limit (URL). For example, a transmitter with a 100 psi URL might have an overpressure limit of 1500 psi.
- Importance:
- Prevents Damage: Exceeding this limit can physically deform or rupture the sensor diaphragm, rendering the transmitter useless.
- Ensures Safety: In high-pressure applications, a ruptured sensor could lead to a dangerous release of process fluid.
- Design Consideration: When selecting an instrument, you must consider potential upset conditions like a blocked discharge or valve failure that could subject the instrument to pressures far above the normal operating range but still within the overpressure limit.
- Note: Even if pressure exceeds the calibrated span but stays below the overpressure limit, the instrument will simply read at its maximum output (e.g., 20.5 mA) and will return to normal operation once the pressure drops back into its calibrated range.
23. How do you properly vent a pressure transmitter in liquid service?
Proper venting is crucial for accuracy in liquid service, as trapped air or gas is compressible and will lead to spongy, inaccurate, or erratic readings.
- Identify Vent Plugs: Most transmitters have small vent plugs on the high and low-pressure sides of the sensor housing.
- Isolate the Instrument: Isolate the transmitter from the process using the block valves on the manifold.
- Control the Venting: With the instrument still under process pressure, I would place a rag over the vent plug to catch any liquid and *very slowly* and *carefully* crack open the vent plug on the high-pressure side first.
- Bleed Until Liquid is Solid: I would allow fluid to seep out until a solid, bubble-free stream is observed. This indicates all the trapped air has been forced out. Then I would securely tighten the vent plug.
- Repeat for Low-Pressure Side: I would repeat the exact same procedure for the low-pressure side vent plug (if applicable, for DP transmitters).
- Return to Service: Once both sides are vented, I would return the instrument to service by following the correct manifold valve sequence.
24. What is the purpose of a snubber or pulsation dampener?
A pulsation dampener, or snubber, is a device installed in the impulse line to protect pressure instruments from the damaging effects of pressure spikes and pulsations.
- How it Works: It works by introducing a restriction or a tortuous path for the process fluid before it reaches the instrument. This restriction smooths out rapid pressure changes, effectively averaging the pressure seen by the sensor. Common types include porous disc (sintered metal), piston-type, or adjustable needle valve snubbers.
- When to Use One:
- Reciprocating Pumps/Compressors: They are essential on the discharge of positive displacement pumps and compressors, which create strong, repetitive pressure pulses.
- Fast-Acting Valves: They can protect instruments from "water hammer" or hydraulic shock caused by the rapid closing of valves.
- Turbulent Flow: In applications with highly turbulent flow that causes rapid pressure fluctuations.
- Downside: A snubber slows down the instrument's response time. It should only be used when necessary, as it may not be suitable for control loops that require very fast reactions. They can also become clogged in dirty services.
25. You find a pressure transmitter with a blown fuse in the control cabinet. What is your troubleshooting process?
A blown fuse is a symptom, not the root cause. Simply replacing it without investigation will likely result in it blowing again.
- Do Not Immediately Replace: I would not replace the fuse right away. First, I need to find out why it blew.
- Inspect Wiring: I would start by visually inspecting the wiring loop from the fuse terminal block all the way to the transmitter. I am looking for obvious signs of a short circuit, such as:
- Damaged insulation from chafing or heat.
- Wires pinched in a junction box cover or conduit fitting.
- The presence of water or corrosion in a junction box or inside the transmitter housing, which can create a path to ground.
- Isolate and Test: I would disconnect the wires at the transmitter's terminals. Then, using a multimeter in resistance mode, I would check for a short circuit between the positive (+) and negative (-) wires. I would also check the resistance from each wire to ground to look for a ground fault.
- Test the Transmitter: If the field wiring checks out, the fault may be internal to the transmitter. I would connect the transmitter to a separate test power supply on the bench. If it immediately draws excessive current, the transmitter's electronics have failed and it needs to be replaced.
- Rectify and Replace: Once I have found and fixed the root cause of the short circuit (e.g., repaired damaged wiring, dried out a junction box), only then would I replace the fuse with one of the correct type and rating and power up the loop.
One Response